GuiSession


stationCode
uXiterFT_01
scandData
000005098,AS000018REVC023324
ATPName
uXIter_FT_ATP_Full_20241112_093412
startTime
11/14/2024 12:51:47
endTime
11/14/2024 12:51:57
Duration
00:10.622
Status
Fail

Test Results

index TestName Min Limit ResultAnswer Max Limit ResultStatus Duration
0 Initialize Pass 00:00.000
1 init Pass Pass 00:03.906
2 OC :: 12V to 4V DCDC Pass 00:00.000
3 Turn ON 12V source Pass Pass 00:00.036
4 Enable 4V DCDC Pass Pass 00:00.035
5 Delay 1000 Pass Pass 00:01.000
6 Read DUT 4V voltage rail -uX1 3800 3905 4100 Pass 00:00.456
7 Measure the current over 12v current rail 63 Pass 00:00.457
8 CC :: 12V to 4V DCDC Pass 00:00.000
9 Load 4V rail with external resistor Pass Pass 00:00.029
10 Delay 1000 Pass Pass 00:01.000
11 Read DUT 4V voltage rail -uX1 3054 Pass 00:00.453
12 Measure the current over 4v current rail 755 Pass 00:00.451
13 Turn off the external load Pass Pass 00:00.027
14 OC :: 4V to 3.3V LDO Pass 00:00.000
15 Enable 3.3V LDO Pass Pass 00:00.029
16 Delay 1000 Pass Pass 00:01.000
17 Read DUT 3.3V voltage rail 3100 3183 3360 Pass 00:00.466
18 Measure the current over 4v current rail 10 96 70 Fail 00:00.460
Back to List