GuiSession


stationCode
ReaderFT_01
scandData
000005331,AS000016REVC054424
ATPName
Reader2.0_FullATP_SKIP_CAL_20241111_162917
startTime
11/11/2024 14:29:46
endTime
11/11/2024 14:30:56
Duration
01:10.802
Status
Pass

Test Results

index TestName Min Limit ResultAnswer Max Limit ResultStatus Duration
0 Initialize Pass 00:00.000
1 init Pass Pass 00:03.749
2 Reader PWR reset Pass 00:00.000
3 Turn OFF the 4.6V high current source (Vbus4V6H) Pass Pass 00:00.035
4 Release NXP reset line Pass Pass 00:00.035
5 Release all pinout Pass Pass 00:00.027
6 Turn ON the 4.6V high current source (Vbus4V6H) Pass Pass 00:00.033
7 Delay 6000 Pass Pass 00:06.000
8 Reader Connect Pass 00:00.000
9 ConnectToReaderTelnet 192.168.0.151 Pass 00:00.328
10 ReaderSetSerialNumber 000005331 Pass 00:00.322
11 ReaderSetMacAddress 1063A39000EB Pass 00:00.212
12 Restart Reader Pass Pass 00:00.108
13 Delay 6000 Pass Pass 00:06.000
14 ReConnectToReaderTelnet 192.168.0.151 Pass 00:00.324
15 ReaderGetMacAddress 1063A39000EB Pass 00:00.221
16 SendReaderConfigFile Pass Pass 00:06.960
17 Get Main CPU version 2.3.3+1031 Pass 00:00.104
18 Get TX CPU version 0.17 Pass 00:00.730
19 Get BLE0 CPU version 0.10 Pass 00:00.731
20 Get BLE1 CPU version 0.10 Pass 00:00.112
21 Get BLE2 CPU version 0.10 Pass 00:00.725
22 Start TX (warming up) Pass 00:00.000
23 TesterMCU_Set_ISM_Mux Pass Pass 00:00.022
24 Set TX on pwr = 240 freq = 902000 debug Pass Pass 00:01.685
25 Delay 10000 Pass Pass 00:10.000
26 TX Test Pass 00:00.000
27 SetSpectrum Pass Pass 00:00.000
28 Get Last Spectrum Mesurment Freq 902000618.38 Pass 00:00.468
29 Get Last Spectrum Mesurment Amp 20 30.03 100 Pass 00:00.000
30 Set TX on pwr = 240 freq = 928000 debug Pass Pass 00:01.686
31 SetSpectrum Pass Pass 00:00.000
32 Get Last Spectrum Mesurment Freq 928000619.10 Pass 00:00.492
33 Get Last Spectrum Mesurment Amp 29.40 Pass 00:00.000
34 Set TX on pwr = 240 freq = 865700 debug Pass Pass 00:01.675
35 SetSpectrum Pass Pass 00:00.000
36 Get Last Spectrum Mesurment Freq 865700498.74 Pass 00:00.499
37 Get Last Spectrum Mesurment Amp 29.91 Pass 00:00.000
38 TX Test - post calibration Pass 00:00.000
39 Restart Reader Pass Pass 00:00.105
40 Delay 6000 Pass Pass 00:06.000
41 ReConnectToReaderTelnet 192.168.0.151 Pass 00:00.336
42 SendReaderConfigFile Pass Pass 00:09.186
43 SetSpectrum Pass Pass 00:00.000
44 Set TX on 30 928000 open Pass Pass 00:01.684
45 Get Last Spectrum Mesurment Freq 928996000 928000619.10 928004000 Pass 00:00.466
46 Get Last Spectrum Mesurment Amp 28 29.16 32 Pass 00:00.000
47 SetSpectrum Pass Pass 00:00.000
48 Set TX on 25 865700 open Pass Pass 00:01.688
49 Get Last Spectrum Mesurment Freq 865796000 865700498.74 865704000 Pass 00:00.490
50 Get Last Spectrum Mesurment Amp 23 26.17 27 Pass 00:00.000
51 SetSpectrum Pass Pass 00:00.000
52 Set TX on 30 865700 open Pass Pass 00:01.678
53 Get Last Spectrum Mesurment Freq 865796000 865700498.74 865704000 Pass 00:00.505
54 Get Last Spectrum Mesurment Amp 28 29.88 32 Pass 00:00.000
55 SetSpectrum Pass Pass 00:00.000
56 Set TX on 30 865700 open Pass Pass 00:01.680
57 Get Last Spectrum Mesurment Freq 902000618.38 Pass 00:00.486
58 Get Last Spectrum Mesurment Amp 29.71 Pass 00:00.000
59 Set TX off Pass Pass 00:00.307
60 Finalize Pass 00:00.000
61 Finalize Pass Pass 00:00.066